FCPS Part 1 Biostatistics and epidemiologyGynae & Obs Paper 2

Principles of imaging

Principles of ultrasound, Doppler, X-ray, CT, MRI, radioisotopes and lasers

Every imaging method sends energy into the body and reads how tissues reflect, absorb or re-emit it. CPSP asks the basic physics behind each method, why one frequency, voltage or sequence is chosen over another, and how tissues appear as a result, especially in the pelvis and pregnancy.

One of the most tested topics in past papers

Ultrasound

  • Ultrasound is sound above 20 kHz (human hearing is 20 Hz to 20 kHz); diagnostic probes use about 2–15 MHz.
  • The probe's piezoelectric crystal vibrates when voltage is applied (transmit) and produces voltage when echoes strike it (receive).
  • Echoes arise at boundaries of different acoustic impedance (density × speed of sound). Tissue–air reflects about 99.9%, so coupling gel is used and bowel gas blocks the view; a full bladder displaces gas and gives an acoustic window for transabdominal pelvic scans.
  • Depth is calculated from echo time, assuming soft-tissue speed of 1540 m/s (depth = speed × time ÷ 2).
  • Wavelength = speed ÷ frequency. Higher frequency means shorter wavelength and better resolution but less penetration, because attenuation rises with frequency.
UseTypical frequency
Transabdominal obstetric3–5 MHz (lower for obese women)
Transvaginal (probe close to target)5–9 MHz
Breast, thyroid, superficial10–15 MHz
  • M-mode (motion against time on one line) records embryonic heart rate at low acoustic output.
  • Artefacts: acoustic shadowing behind stones, bone and calcification (strong reflection and absorption); posterior acoustic enhancement behind cysts (fluid attenuates little); reverberation (comet tail) between strong reflectors.
  • Transvaginal scanning is the best test to localise the placenta in suspected placenta praevia.

Doppler

  • Sound reflected from moving red cells changes frequency in proportion to their velocity and the cosine of the beam angle (Doppler shift).
  • Colour Doppler: red = flow towards the probe, blue = away (nothing to do with oxygenation); mosaic = turbulence or aliasing.
  • Pulsed Doppler samples a chosen depth but aliases when the shift exceeds half the pulse repetition frequency (Nyquist limit); fix by raising the scale or lowering the baseline. Continuous-wave Doppler (two crystals, as in hand-held fetal heart detectors) has no depth information and does not alias.
  • Indices: pulsatility index = (S − D) ÷ mean; resistance index = (S − D) ÷ S. As ratios, the angle cancels out, so they are used for umbilical and uterine arteries. Absent or reversed umbilical end-diastolic flow means high placental resistance.
  • Pulsed Doppler has the highest thermal output; use it sparingly in the first trimester.

X-rays

  • Electrons from a heated cathode strike a tungsten anode, producing mainly bremsstrahlung (braking radiation); about 99% of the energy becomes heat.
  • kV controls photon energy and penetration; mAs controls photon number (and dose). High kV lowers contrast.
  • Photoelectric absorption (photon fully absorbed, probability proportional to Z³) creates contrast: calcium, iodine (Z 53), barium look white. Compton scatter dominates in soft tissue, fogs the image (reduced by grids) and is the main source of staff dose.
  • Mammography uses low kV (25–30 kV) to maximise photoelectric soft-tissue contrast.

Computed tomography

  • A rotating X-ray beam takes many projections; the computer calculates attenuation in each voxel (filtered back-projection or iterative reconstruction), removing superimposition.
  • Hounsfield units: water 0, air −1000, fat about −100, soft tissue +20 to +60, fresh blood 40–70, bone up to +1000. Fat at −100 HU in an ovarian mass means a dermoid.
  • Windowing: level is the centre HU, width the range shown.
  • CT carries a much higher dose than plain films.

Magnetic resonance imaging

  • Hydrogen protons align in a strong field (1.5–3 T) and precess at the Larmor frequency (about 42.6 MHz per tesla). A radiofrequency pulse at that frequency tips them (resonance); as they relax they emit signal. No ionising radiation.
  • T1 (spin–lattice) and T2 (spin–spin) relaxation create contrast:
TissueT1-weightedT2-weighted
Water, CSF, urine, cystsDarkBright
FatBright (drops with fat suppression)Bright
Subacute blood (methaemoglobin)Bright (stays bright with fat suppression)Variable
Haemosiderin, ironDarkDark (blooms on gradient echo)
  • Endometrioma stays T1 bright with fat suppression; a dermoid loses signal.
  • Gadolinium contrast shortens T1 (enhancing tissue turns bright).
  • Hazards: the always-on static field pulls ferromagnetic objects; RF pulses cause heating; gradients cause noise. Pacemakers and some implants are contraindications.

Radioisotopes

  • A tracer emits gamma rays detected by a gamma camera (collimator, sodium iodide crystal, photomultipliers).
  • Technetium-99m: ideal: half-life 6 hours, pure gamma of 140 keV, no beta (low patient dose), from a molybdenum-99 generator.
  • PET: fluorine-18 FDG emits positrons (paired 511 keV photons) and tracks glucose use.

Lasers

  • Light Amplification by Stimulated Emission of Radiation: needs a population inversion; light is monochromatic, coherent and collimated, giving high power density when focused.
LaserWavelengthBehaviour and use
CO210,600 nm (far infrared)Strongly absorbed by water: shallow, precise vaporisation (CIN, vulval and vaginal lesions); cannot work through fluid
Nd:YAG1064 nm (near infrared)Passes through water and clear fluid: deep coagulation; fetoscopic ablation of placental vessels in twin–twin transfusion
ArgonVisible blue-greenAbsorbed by haemoglobin and pigment: retinal photocoagulation

Remember Higher ultrasound frequency sees finer detail but not as deep; that is why the transvaginal probe gives the sharper picture.

Exam trap Red and blue on colour Doppler show direction relative to the probe, not arterial and venous blood.

Exam trap Bright on T1 is not always fat: blood stays bright when fat suppression is applied.

Test yourself on this topic

5 of the 81 questions in our bank on this topic. Choose an answer to see why every option is right or wrong.

Question 1Easy

A gynaecologist uses a laser at colposcopy. The light produced is a single wavelength, coherent and tightly collimated. Which physical process generates this light?

Question 2Easy

An unenhanced CT shows a thin-walled adnexal cyst whose contents measure 5 Hounsfield units, which the radiologist describes as simple fluid. On what basis is the Hounsfield scale set?

Question 3Medium

For an abdominal radiograph in a heavily built woman, the radiographer raises the tube voltage from 70 kV to 85 kV. What change in the X-ray beam does this produce?

Question 4Easy

In a woman with secondary infertility, a hysterosalpingogram outlines the uterine cavity and shows spill from both tubes. Why does the injected contrast medium appear white on the X-ray?

Question 5Medium

A midwife uses a hand-held fetal heart detector at 16 weeks. It gives an audible heartbeat but no image and no information about depth. Which technique does the device use?

Practise all 81 questions on this topic

Where this sits in the CPSP syllabus

  • Obstetrics & Gynaecology, Paper 2: Biophysics, Principles of ultrasound, Doppler, X-ray, CT, MRI, radioisotopes and lasers